Running Head: Ambulatory Oxygen for Exercise-Induced Desaturation & Dyspnea in COPD
Funding Support: No specific, designated funds were received for this study. However, during the term of this study, Dr. Turner received an unrestricted grant from the Linde Real Fund to conduct a clinical trial of ambulatory oxygen. AUTHOR NOTE: Funding support section should speak specifically to funding for the study discussed in the article—please review adjusted sentences.
Date of Acceptance: October 21, 2015
Abbreviations: long-term oxygen therapy, LTOT; chronic obstructive pulmonary disease, COPD; 6-minute walking test, 6MWT; endurance shuttle walk test, ESWT; exercise-induced desaturation, EID; quality of life, QoL; randomized controlled trials, RCTs; pulmonary rehabilitation, PR; standard deviation, SD; activities of daily living, ADLs; baseline/transitional dyspnea scores, BDI/TDI; blood oxygen saturation, SpO2; Chronic Respiratory Questionnaire, CRQ; Hamilton Depression and Anxiety Scale, HADS; pulmonary function tests, PFTs; minimal clinically important difference, MCID; St. George’s Respiratory Questionnaire, SGRQ; work rate, WR; partial pressure of oxygen in arterial blood, PaO2; 5-minute walk test, 5MWT; American Thoracic Society, ATS; demand oxygen devices, DODs; pendant reservoir cannulae, PRC; metabolic equivalents, METS; not reported, NR; confidence interval, CI; oxygen conserving devices, OCDs; continuous flow nasal cannulae, CFNC
Citation: Ejiofor SI, Bayliss S, Gassamma B, Turner AM. Ambulatory oxygen for exercise-induced desaturation and dyspnea in chronic obstructive pulmonary disease (COPD): systematic review and meta-analysis. Chronic Obstr Pulm Dis. 2016; 3(1): 419-434. doi: http://doi.org/10.15326/jcopdf.3.1.2015.0146
Background
Chronic obstructive pulmonary disease (COPD) is a leading cause of morbidity and mortality, with prevalence anticipated to increase due to persistent exposure to COPD risk factors and an ageing population.1 Hypoxaemia and dyspnea are hallmark features; in many individuals gaseous exchange becomes more impaired as the disease progresses, leading to hypoxemia at rest or on exercise alone. Several mechanisms causing this are diffusion limitation, ventilation/perfusion mismatches and shunting. Exercise-induced desaturation (EID) is likely to contribute to exercise limitation in COPD. EID is defined as an arterial oxygen saturation <88% (nadir) measured by pulse oximetry during exercise.2 Current evidence suggests that EID is common, and may lead to worse quality of life(QoL) and increased mortality.3,4
Ambulatory oxygen therapy is defined as the use of supplementary oxygen during activities of daily living. Although the provision of ambulatory oxygen varies widely across countries, its use is generally accepted in patients who exhibit exertional desaturation if there is a demonstrable improvement in exercise capacity or dyspnea scores.5,6 While the benefits of oxygen treatment in the form of long term oxygen therapy (LTOT)7,8 are well recognized, the long term benefits of ambulatory oxygen in COPD are controversial. A Cochrane review demonstrated benefit of ambulatory oxygen in improving exercise capacity in moderate to severe COPD, albeit mostly in single assessment studies.9 In a subsequent review evaluating ambulatory oxygen in the context of pulmonary rehabilitation programs, no benefit was demonstrated.10
Current clinical use and associated costs of ambulatory oxygen suggest that assessing longer term benefits of ambulatory oxygen in COPD in a domiciliary and rehabilitation setting is required. Thus the aim of this review was to determine:
- The clinical effectiveness of long term ambulatory oxygen therapy (defined by more than 6 weeks duration) in patients with COPD exhibiting EID or exertional dyspnea who do not meet the criteria for LTOT;
- The relative clinical effectiveness of different methods of delivering ambulatory oxygen in COPD patients and the optimum mode of delivery of the intervention;
- The cost-effectiveness of ambulatory oxygen in COPD patients not on LTOT compared to usual care and the relative cost-effectiveness of different methods of delivering ambulatory oxygen.
Methods
Studies were eligible for inclusion if they included adult patients (18 years and over) and met the criteria in Table 1: Randomized controlled trials (RCTs), non RCTs and crossover studies were included for clinical and cost effectiveness aims. To address aim 2: ambulatory oxygen delivery device efficacy, single assessment studies and patients who met LTOT criteria were included (for this aim only). The definition of COPD was broadened to include emphysema and chronic bronchitis. The protocol is registered with Prospero (CRD42012002984).
Search Strategy
To maximize efficiency, 2 search strategies were undertaken, under the guidance of an information search specialist, as outlined in the supplementary methods. The literature search included the following databases; MEDLINE, MEDLINE in Process, EMBASE, Cochrane-(CENTRAL), Science Citation Index, National Health Service Economic Evaluation Database, PEDro, Health Technology Assessment database, ClinicalTrials.gov, Cumulative Index to Nursing and Allied Health Literature, and the Cochrane Airways specialized register. Following this, the references of each included study, as well as any review articles found, were searched for additional articles that may contain further studies.
Database searches were conducted between 1980 and June 2013 inclusive as older studies; pre 1980 would have been before the introduction of LTOT. An updated search run to June 2014 yielded no further articles suitable for inclusion.
Study Selection, Data Extraction and Risk of Bias
Two reviewer authors (SE and AT) independently selected trials for inclusion in the review. They first reviewed titles and abstracts and obtained full articles judged as potentially eligible. These were screened independently and any disagreements resolved by discussion or referral to a third reviewer. Methodological quality was assessed by Jadad Scale.11 For comparative studies we also assessed adequacy of sequence generation, allocation concealment and blinding.
Data relating to all aspects of the study was extracted by 1 reviewer using a standardized, piloted data extraction form in Access (Microsoft); a second reviewer checked all numerical data. Data was extracted on study and population characteristics, intervention and comparator characteristics, study quality and results.
Statistical Analysis
All trial data was combined using RevMan 5.2 (Cochrane Collaboration). Studies consisted of parallel RCTs, randomized crossover trials and randomized single assessment studies. Analysis of studies was dichotomized to those which involved structured pulmonary rehabilitation (PR) and those which were domiciliary. Only the results from the same exercise test protocols were combined. Pre- and post-test exercise outcomes reported for both intervention groups were from tests performed on compressed air.
Data was pooled with the generic inverse variance method using change from baseline with the appropriate means and standard deviations for each study if provided. For studies that reported pre- and post-training means and respective standard deviations, the change in outcome measure pre-training means were subtracted from the post-training means and standard deviations (SD) for pre- and post-training were pooled.
A fixed effect was used in all analyses unless heterogeneity (defined as I2>30%) was present in which case a random effect was used. The meta-analyses are presented using forest plots from RevMan. Data not in a format suitable for meta-analysis have been reported in narrative form.
Results
Description of Studies
Twenty-three studies (620 patients) were included12-34: 9 studies evaluated the clinical effectiveness of ambulatory oxygen and 14 evaluated the impact of the delivery device. Generally studies were small with primary outcomes focused on exercise capacity, Borg dyspnea scores and exercise saturations.
Figure 1 summarizes the trial flow in the PRISMA format for reporting systematic reviews. Tables 2 and 3 detail the characteristics of included studies.
Methodological Quality of Included Studies
The quality of studies assessed by the Jadad scale was poor in 12 studies (score<2), good in 9 (score 3 or 4) and excellent in 2 (score 5) (Supplementary Table 1).
Clinical Effectiveness of Ambulatory Oxygen
Exercise Capacity
Four RCTs12,13,17,18 used 6MWT distance in assessing exercise capacity. In the PR studies13,17 the common effect (weighted mean difference) was 21.97 meters in favor of the control group. This result was mirrored in the meta-analysis of the domiciliary studies12,18; common effect 0.56 meters in favor of control. However, neither result was statistically significant (95% confidence interval [CI] -77.25 to 33.32) and (95% CI -26.96 to 28.08) respectively, Figure 2. Two PR RCTs19,20 used ESWT; they differed substantially in their results, hence the pooled figure may be less meaningful. It was not statistically significant with a common effect of 150.02 seconds in favor of oxygen (95% CI -124.74 to 427.77 seconds, Figure 2). Two PR RCTs 13, 15 used incremental work rate exercise testing via cycle ergometry to assess exercise capacity and found no difference with oxygen; 1.45 watts in favor of control (95% CI -13.54 to 10.64 watts), Figure 2. Two PR RCTs13,15 measured constant power exercise tests using cycle ergometry. There was a strong trend toward longer exercise time with oxygen (2.76 minutes longer [95% CI -0.07 to 5.58 minutes] Figure 2).
Two studies (both domiciliary)14, 16 could not be used in the meta-analysis as 1 study provided p values only 14 and the other was the only one to use 5MWT.16 Neither showed any statistical benefit of long term ambulatory oxygen on 6MWT distance (p=0.9) or 5MWT distance for individual participants.
Borg Dyspnea Scores
End of test Borg scores were measured in 2 of the meta-analysed PR studies that used 6MWT13,17 and 2 RCTs using incremental work exercise testing.13,15 Heterogeneity between the studies was not significant (I2=0%). Neither were statistically significant; effect 0.16 (-0.61 to 0.93) and 0.25 (-0.81 to 1.31) respectively; (Figure 3).
Quality of Life
This was measured by the Chronic Respiratory Disease Questionnaire (CRQ) in 7 studies 12-16,18,19 3 of which were PR studies. In the PR studies there was no statistical difference between oxygen and control in any domain, however there was a strong trend to statistical significance in CRQ emotion and fatigue favoring oxygen; p=0.06 and p=0.09 respectively (Figure 4). In the domiciliary studies the effect in all domains was in favor of oxygen treatment, significant only in dyspnea (p=0.002) and fatigue (p=0.01). (Figure 5)
Impact of Device on Ambulatory Oxygen Delivery
Fourteen studies evaluated a variety of oxygen conserving devices (OCDs) and cylinder types21-34 (Table 3). Modes of delivery included demand oxygen devices (DODs), pendant reserve cannulae (PRC) and continuous flow nasal cannulae (CFNC). In some studies, multiple OCDs were compared. Where this was the case, the most effective OCD was used in the meta-analysis comparing it to standard CFNC.
Five OCD studies22,24,26,30,34 were used in 2 separate meta-analyses. Tests for heterogeneity were not significant (both I2=0).
Exercise Capacity
Three of the 5 meta-analysed studies compared DODs against standard CFNC using 6MWT as an outcome measure. 26,30,34 DODs and CFNC did not differ (CI-10.28 to 1.60), (Figure 6).
Exercise Oxygen Saturations
Oxygen saturations were used in 2 of the 5 studies 22,24 with methods of delivery being PRC and DODs respectively. Oxygen saturations were 2.03% higher with OCDs compared to CFNC, (95% CI 0.09 to 3.98, p=0.04, Figure 6). In contrast, Roberts et al demonstrated that DODs caused a significantly longer period of desaturation <90% than CFNC p<0.001.
Oxygen Utilization
Three studies21,23,28 evaluated the ability of OCDs versus standard CFNC to conserve oxygen in patients during exercise by calculating an oxygen conservation ratio at 90% blood oxygen saturation (SpO2) (i.e., the relative difference in oxygen flow required to maintain oxygen saturation with both devices at 90% SpO2). Two studies used DODs23,28 and 1 PRC.21 Although a higher DODs setting was required to maintain equivalent oxygen saturations, DODs still conferred 2.9-7.6 fold oxygen savings which was significant when compared to standard CFNC p<0.001. A similar result was demonstrated with PRC; oxygen utilization was significantly lower than CFNC p<0.05.21
Refillable Cylinders and Home Activity
Individual studies demonstrated no significant difference in exercise capacity with either refilled cylinders or portable oxygen concentrators when compared to standard commercial cylinders.27,31 Patterns of either home activity or oxygen utilization were not significantly different when lightweight cylinders were used compared to heavier E cylinders.33
Cost Effectiveness
There were no results from the searches assessing cost incremental ratios or cost effectiveness for ambulatory oxygen in addition to usual care.
Discussion
This review evaluated the use of ambulatory oxygen in 3 separate ways: efficacy against placebo, optimum mode of delivery and cost. First, we have shown that ambulatory oxygen is unlikely to be effective at improving functional exercise capacity or symptoms of breathlessness in unselected patients with EID, although there may be some patients who respond well. Second, we have shown that use of OCDs is appropriate, in that outcomes are similar or better to CFNC but at lower oxygen consumption. Since the review showed no significant overall effect of oxygen, and there were no published cost-effectiveness studies, we were unable to calculate an incremental cost effectiveness ratio. Each area of the results will be discussed here in turn.
Clinical Effectiveness of Ambulatory Oxygen
The prescription of ambulatory oxygen therapy in COPD is governed by an improvement in exercise capacity or Borg dyspnea scores, wherein a 10% improvement in distance walked or reduction ≥ 1 in Borg Score indicates it should be used in patients with EID > 4%.5 The majority of our results suggest that there is not a benefit of this magnitude (Figures 2 and 3). It is important to highlight that in the included studies the definition of EID was not uniform, ranging from mild (<92%) to more severe desaturation (<88%). Since results for some outcomes were heterogeneous it is important to consider how patient selection influenced results. Two studies12,18 used exertional dyspnea as their main inclusion criterion and were analyzed separately, although 1/3 of patients in the Moore et al study also exhibited EID. Few studies specified acute responders (i.e., those with >10% improvement in walking distance). The only study19 which did specify, used ESWT as the outcome measure and showed significant improvement in those randomized to ambulatory oxygen compared to compressed air, implying that appropriate patient selection is important in optimizing clinical outcome. Interestingly however, 7 studies12-16,18,20 included a single assessment acute oxygen test for all participants as part of their protocol. All but one20 demonstrated a significant mean improvement in exercise capacity for participants with acute oxygen therapy compared to compressed air which is in line with a previous Cochrane review.9 Whether the acute effects of oxygen persist in the long term in trained or untrained participants was addressed in 2 of the included studies13,14 as they all included a repeat acute oxygen test at the end of the study period. Rooyackers et al13 and Eaton et al14 both demonstrated that this acute effect was no longer statistically significant when using 6MWT as the outcome measure. It would therefore seem that any acute improvement in walking distance observed in a single assessment study is not maintained. This has implications for clinical practice as prescriptions for ambulatory oxygen based on improvements in exercise capacity observed in single assessment studies as suggested by guidelines5 are not sustained or are minimal in the longer term.
The way in which exercise capacity was measured may also be relevant to the interpretation of results. Largely it was measured using field tests (ESWT or 6MWT) which, particularly in the case of the 6MWT, have been suggested to relate more to activities of daily living.35,36 The meta-analysis of exercise capacity with respect to 6MWT slightly favored the control group and showed no benefit of ambulatory oxygen, implying that in normal daily life it would not be expected to help patients be more active. In the ESWT meta-analysis, although the result was not statistically significant in favor of oxygen therapy, it may be clinically significant. The minimal clinically important difference (MCID) for the ESWT is estimated at 45-85 seconds.37 The common effect of 150 seconds exceeded this although this has to be approached with caution because the heterogeneity between the 2 studies19,20 was considerable. Both studies enrolled patients exhibiting EID>4% and predominantly severe COPD. During the ESWT in the Dyer et al study, patients had their oxygen cylinders carried by an assistant whereas this was not the case in Ringbaek et al. It is recognized that any improvement in exercise capacity can be negated by the weight of the ambulatory oxygen system if carried by patients alone,26,38 thus it is possible the Dyer study overinflated the benefits of treatment in real life by providing assistance. The meta-analysis of endurance cycle time strongly favored oxygen treatment and was close to approaching statistical significance. Furthermore, this is likely to be an underestimate of the effect of ambulatory oxygen treatment as 16 endurance tests in the oxygen group were terminated at 30 minutes compared to 6 endurance tests in the control group.15The responsiveness of constant work rate tests such as the ESWT and cycle endurance to intervention is far better than the 6MWT39 which may explain the discrepancy between these different exercise outcome results. However, at present no MCID for cycle ergometry has been established and its relationship to activities of daily living is unclear.
Some studies13,15,17,19,20 were carried out as part of pulmonary rehabilitation. Although no statistical benefit of long term ambulatory oxygen was demonstrated, exercise capacity (particularly 6MWT) in the pulmonary rehabilitation studies exceeded that of the domiciliary studies12,18 whether patients were randomized to ambulatory oxygen or placebo, Figures 2.1 and 2.2 What is also clear is that the improvement in 6MWT distance gained by pulmonary rehabilitation far exceeds that gained by ambulatory oxygen.13 This supports current guidelines5 in that any assessment of ambulatory oxygen therapy should be made following pulmonary rehabilitation.
Finally, it is important to consider the meta-analyses possible using CRQ domains. In general, there was a benefit of ambulatory oxygen on QoL, which was most marked in the dyspnea and fatigue domains; the others were not statistically significant but showed a strong trend in the same direction. However, the effect size was only 0.28 and 0.17 in the statistically significant domains, which is less than the 0.5 points considered to represent the MCID for an individual domain.40 This explains the apparent discrepancy between Borg score and CRQ dyspnea score meta-analyses; neither were clinically significant, hence they concur, despite the fact that statistically the CRQ score was better with oxygen. The effect size is also considerably less than the effect on CRQ of pulmonary rehabilitation (0.62 points)41 confirming that assessment for ambulatory oxygen should only be done after rehabilitation.
Impact of Device on Ambulatory Oxygen Delivery
The mode of delivery of ambulatory oxygen therapy largely encompasses the cylinder type and the interface used. Oxygen conserving devices (DODs/PRC) have been used for many years and the theory supporting their use has been discussed elsewhere.23 In short classic DODs, devices deliver oxygen during the initial phase of inspiration alone while PRCs store oxygen during exhalation to be delivered at the next inspiratory phase. The role of OCDs is to conserve oxygen however, in doing so they must perform equivalently to standard CFNC in ameliorating oxygen desaturation. Some studies 22,24,26,30,34 included in the meta-analysis indicate that oxygen conserving devices perform equivalently to standard CFNC at maintaining exercise oxygen saturation and 6MWT. Importantly, however, it is likely that oxygen conserving devices achieve these outcomes at a much lower oxygen utilization than CFNC. This is likely to have implications in terms of oxygen cylinder usage and subsequent cost. Although the primary role of OCDs is not to improve exercise performance over CFNC, patients use these devices on the premise that they are ambulatory. Thus, it is reassuring that functional performance is not negated by the weight of the system.
It is important to note that there is considerable variation between OCDs and their performance and in the meta-analyses the best performing OCD was used. Notably in the study by Palwai et al, one of the DOD systems performed no better than ambulation on room air. Also there may be some patients who do not tolerate OCDs and their use may cause further desaturation. This is supported by the study by Marti et al34 in which use of DODs failed to correct EID in approximately 20% of participants. This has been explained by changing respiratory pattern on more strenuous exertion where by the predominance of mouth breathing rather than nasal fails to trigger the flow sensor.26 Oxygen cylinders have historically been the main method of delivering portable oxygen, however, newer devices such as portable concentrators are now available. These devices have variable weight with cylinders weighing between 3.4-8kg and portable concentrators between 3-4kg. As mentioned previously, the weight of the ambulatory oxygen system (cylinder in this case) is an important factor as the portable oxygen system should be carried by the patient during their ambulatory oxygen assessment and, following this, in their ambulatory activity. The weight of the portable oxygen system has previously been cited as one of the reasons why patients with COPD have poor concordance with ambulatory oxygen prescriptions.42 An interesting observation from 1 of the studies33 is that ambulatory cylinder weight had no impact on home activity. One would hypothesize that a lower weight would facilitate ambulation. This was not the case despite prior efforts of focused education on increasing oxygen treatment understanding and ambulation.
Evidence from the included studies suggests that refilled oxygen cylinders/portable oxygen concentrators perform as well as standard commercial cylinders.
Limitations
The quality of data in this systematic review was poor to average as limited by small sample sizes and high risk of selection and detection bias, especially in the older studies. In addition, patient selection and heterogeneity of outcomes means that interpretation of the results should be guarded. Outcomes in this review did not include other clinically relevant endpoints such as exacerbations or mortality. Finally, there were no studies that assessed cost effectiveness. As clinical effectiveness could not be proven in any domain it was not possible to answer whether ambulatory oxygen is cost effective.
Conclusions
This review suggests that long term use of ambulatory oxygen has limited benefit compared to ambient air when assessed by functional exercise capacity or Borg dyspnea scores. It may, however, have greater efficacy in improving exercise performance when assessed by endurance tests. The impact on quality of life, although statistically significant, is unlikely to be clinically relevant. Where ambulatory oxygen is used, oxygen conserving devices should be used in preference to standard nasal cannula as they generally perform equivalently but at a lower oxygen utilization.
Acknowledgements
The authors would like to thank Dr. David Moore for assistance in designing the protocol.
SE was the primary reviewer, conducted all statistical analysis and drafted the paper. AMT was second reviewer and reviewed the manuscript. SB conducted the literature searches. AG contributed to protocol design.
Declaration of Interest
SE is funded by Linde Real Fund and a non -commercial grant from CSL Behring. AG is funded by the West Midlands Chest Fund. AMT has received grants during the term of this work from the Linde Real fund, the NIHR andAlpha-1 Foundation. The most relevant of these for the submitted article is the Linde Real fund which was an unrestricted grant to conduct a clinical trial of ambulatory oxygen. She has also received payments or benefits of <£10000 over the last 5 years from several manufacturers of respiratory medicines, including GlaxoSmithKline, Novartis, Astra-Zeneca, Pfizer and Boehringer for educational talks and travel to meetings. The above funders had no role in data collection, statistical analysis or writing of this manuscript. SB has nothing to declare.